TRUFORMA® requires minimal routine maintenance. As part of standard preventative maintenance, the air filter located on the rear of the instrument should be replaced monthly to ensure optimal performance.

To replace the air filter:

  • Power off the instrument and unplug the power cable from the rear
  • Use a flat-head screwdriver to remove the filter retainer (note the orientation - an arrow is moulded into the plastic)
  • Remove and discard the used filter
  • Insert a new filter (the filter is symmetrical and has no front or back)
  • Refit the retainer, reconnect the power cable, and power the unit back on